BGD 865 is a small, simple and economic xenon test chamber. It uses a low power air-cooling xenon lamp to produce enough big irradiance energy in a small space. Moreover, through a special catoptrical system to ensure every exposure sample get the homogeneous irradiance
distribution.
BGD 865 is equipped with extended UV filters, thus permit UV which is below the normal cut-on of natural sunlight to pass equal to simulate sunlight which doesn't come through atmosphere .They are used to produce faster or more severe test results. This test is called as "Artificial Accelerated Weathering Test".
Operator can set all required test parameters (Irradiance, test time, BPT etc) through the touch screen, and can check its running status at any time. All running parameters can be exported to computer through the USB interface.
Click Here to Download BrochureMain Technical Parameters:
Xenon Lamp | a 1.8KW imported originally Xenon lamp or a Chinese 1.8 KW Xenon lamp |
---|---|
Fliter | Extended UV filter (Also can choose daylight filter oe window glass filter) |
Irradiance monitor methods | 340nm or 420nm(Narrow band) or 300nm ~ 400nm (Broad band) (Choose anyone before ordering) |
Lamp Lifetime | Near 1,500 hours |
Exposure Area | 1,000 cm2 (can put 9 standard samples 150×70mm) |
Adjustable Irradiance Range | 0.35~0.70 W/m2 (Single point of control: 340nm) 0.5~2.1 W/m2 (Single point of control: 420nm) 30 W/m2~100 W/m2 (Full spectrum :from 300~400nm) |
Adjustable Black Board Temperature Range | RT+30oC~80oC |
Interior Material of Chamber | Stainless steel -SUS 304 material |
Exterior Material of Chamber | Powder coating |
Overall Size | 1050×530×375mm |
Weight | 65 KG |
Power Supply | 220V 50/60HZ;Max Current 13A |
Ordering Information:
BGD 865 | Bench Xenon Test Chamber(Equipped with Chinese lamp) |
---|---|
BGD 865/A | Bench Xenon Test Chamber(Equipped with American lamp) |
BGD 8155 | Xenon Lamps (1.8 KW) |
BGD 8156 | American Xenon Lamps (1.8 KW) |
BGD 8140 | Xenon Calibration Radiometer (340nm) |
BGD 8141 | Xenon Calibration Radiometer (420nm) |
BGD 8142 | Xenon Calibration Radiometer (300nm~400nm) |