C306H Water Vapor Transmission Rate Test System is designed and manufactured based on infrared sensor method and conforms to the requirements of ASTM F1249 and ISO 15106-2. This instrument can be used to measure the water vapor transmission rate of barrier materials with high and medium moisture barrier properties with a wide testing range and high testing efficiency. C306H is applicable to determination of water vapor permeability of plastic films, sheeting, paper, packages and other relative packaging materials in food, pharmaceutical, medical apparatus, consumer goods, photovoltaic and electronic industries, etc.
Infrared sensor
Accurate data
High Efficiency
Intelligent control
Safe and reliable
Space saving
Powerful functions
ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Parameters
Parameters/Model | C306H | |
---|---|---|
Test range | g/(m2·day) (Standard area 50cm2) | 0.02~40 |
g/(m2·day) (MASK area 5cm2) | 0.2~400(Optional) | |
g/(m2·day) (MASK area 1cm2) | 1~2000(Optional) | |
Resolution | g/(m2·day) | 0.0001 |
Repeatability | g/(m2·day) | 0.02 or 2%, take the greater |
Test Temperature | oC | 15~50, 5~60(Optional) |
Temperature fluctuation | oC | ±0.15 |
Test Humidity | %RH(Within standard temperature range) | 100%, 5~90%±2% |
Additional Functions | DataShieldTM note3 | Optional |
GMP Computer System requirement | Optional | |
CFR21Part11 | Optional |
Technical Specifications
Test Cell | 6 Cells |
---|---|
Specimen Size | 4.6" x 4.6"(11.7cm x 11.7cm) |
Specimen Thickness | <=120 Mil(3mm) |
Standard test area | 50cm2 |
Carrier Gas | 99.999% High purity nitrogen(Outside of supply scope) |
Carrier Gas Pressure | >=40.6 PSI / 280 kPa |
Port Size | 1/8" Metal tube |
Instrument Dimension | 23.6" H x 19.2" W x 25.9" D (60cm x 49cm x 66cm) |
Power Supply | 120VAC±10% 60Hz / 220VAC±10% 50Hz(one of two) |
Net Weight | 220Lbs(100kg) |
Product Configuration
Standard Configuration | Instrument mainframe, tablet, sampler, vacuum grease, Φ6 mm PU tubing |
---|---|
Optional Parts | Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
Note | The gas supply port of the instrument is Φ6 mm PU tubing(pressure>=79.7 PSI / 550 kPa), customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.