Medical face masks flame retardant tester is to determine the fire resistance performance of medical face masks and full face masks. The tester can provide a specific flame and make the dummy head with facepiece move to the flame under a specific speed. The operator can observe the effects of the flame on the masks and facepiece. The tester is suitable for medical face mask, respiratory protective equipment and full face masks.
Click Here to Download BrochureEN 149, GB/T 19083, GB 2626, YY 0469
Item | Detail |
---|---|
Moving speed | 60±5mm/s |
Flame height | 40±4mm |
Flame temperature | Thermal couple Up to 850℃ |
Burner | TEKLU burner |
Burner adjust height | 30mm |
Distance between burner and mask | 20±2mm |
Timer | 0-9999.9s ±0.1s |
Power supply | 220 AC 50Hz |
Dimension | 950mm×400mm×850mm 37.4inch×15.75inch×33.46inch |
Net weight | 65 kg 143.3lb |
No. | Product |
---|---|
2042801 | Medical face mask flame retardant tester |
2042802 | TEKLU burner |