Face masks particle filtration efficiency (PFE) tester is test machine to determine the particulate filter penetration and filtration efficiency for respiratory protective devices and face masks. The tester with particle generator can provide sodium chloride particles and paraffin oil particles. It can measure the particle penetration, filtering efficiency and the breathing resistance delta P. It is widely used in quality control, research and development, certification, manufacturing or testing industry, etc.
Click Here to Download BrochureASTM F2299, ASTM F2100, EN 149, EN 13274-7, NIOSH 42 CFR Part 84, YY 0969, YY 0469, GB 2626, GB 19083, GB/T 32610
Item | Detail |
---|---|
Salt particle CMD | 0.075μm |
Oil particle CMD | 0.185μm |
Flow rate | 8-100 L/min ±2.0% |
Particle concentration detection range | (0.001-100)μg/L |
Test pressure range | 0-2500Pa ±2.0% |
Test aerosol | DEHS, Paraffin oil, NaCl |
Test area | 100cm2 |
Particle counting | Photometer |
Display | 10" Touch screen control panel |
Data export | Embedded printer |
Power supply | 220V 50-60 Hz |
Dimensions | 700mm x 600mm x 1520mm 27.56inch x 23.62inch x 59.82inch |
Net weight | CA. 85kg CA. 187.39lb |
No. | Product |
---|---|
2044301 | Face masks particle filtration efficiency tester |
2043302 | Oil aerosol generator |
2043303 | NaCl salt aerosol generator |
2043304 | Photometer |