T446 Respiratory Total Inward Leakage Test Machine

Respiratory total inward leakage test machine is designed to determine total inward leakage of respiratory according to EN 149. NaCl aerosol generator provides 100L/min air flow with particles. The leakage is a measured ratio by mean concentration of breathing zone and challenge concentration of inhalation and exhalation duration. Automatic control and calculated system provides easier and stable operation.

Click Here to Download Brochure   
  • Control cabinet with test chamber.
  • English switch language software.
  • Real-time leakage display and automatic calculation of leakage.
  • Support data storage, report export.

EN 149 Respiratory protective devices – Filtering half masks to protect against particles – Requirements, testing, marking

Specification

Item Detail
Air flow rate 100L/min
NaCl particle concentration 10±2mg/m3
NaCl particle Size distribution is 0.02 to 2μm with a mass median diameter of 0.6μm
Particle measuring range 0.001-200mg/m3, ±1%
Sampling pump 0.50-4L/min
Movement indicator Yes
Power supply 220V 50Hz
Dimension 2400mm x 1200mm x 2370mm
Net weight 210kg